+86 139 2552 1330 [email protected] Dongguan, Guangdong, China
Engineering response within 2 working days
Medical devices

Dispensing actuator for a benchtop analyser

A diagnostic equipment maker required repeatable 2 mm actuation with low acoustic noise and a documented validation trail for their design history file.

Medical devicesCustom design
Round holding electromagnet re-wound for continuous duty, case cover
Actual product and application hardware from the project.

The constraint

Off-the-shelf units exceeded the 50 dB(A) noise limit at the operator position.

What we changed

Custom coil winding and an elastomer end stop to soften the impact.

Result

Measured 44 dB(A) at 1 m across a 50-unit pilot batch.

Client identity withheld at their request. Test conditions and measured values are as recorded in the project file; they apply to the configuration described and should not be read as a general specification.

Have a drawing, a sample, or just a stroke and a force?

Send it over. An engineer reviews every enquiry and replies with a feasible concept, not a brochure.